Differential Pressure Leak Tester AXIS-677 - Measures the Pressure Loss (ΔP) between the test part and the reference volume over a fixed period of time. The leak value is determined from the pressure loss measured during the test cycle. The result is displayed in differential pressure units.
Differential Pressure Loss Measurement: The procedure is based on comparison of differential pressure loss between the test part and reference volume.
Fast 24-bit A/D converter with real-time signal processing: Real-time processing of the measurement signal
Testing Pressure Range: Vacuum to 16 bar
Available Ranges: Vacuum, 0-0.1 bar, 0-1 bar, 0-7 bar, 0-10 bar, 0-16 bar. Pressure ranges above 10 bar are available as customized options.
Differential Measuring Range Pressure Sensor: ±500 Pa, ±1000 Pa, ±2500 Pa, ±5000 Pa
Pressure Regulation: Mechanical or electronic regulator options available depending on pressure range.
Programs: Up to 128 configurable test programs
Display: 5-inch colour touch screen
Leak Port: Calibration Leak Port
Calibration: Auto Zero Calibration
Data logging: Standard USB port
Data storage: Up to 100,000 records
Unit: Pa, Pa/s, mbar, psi, bar, kPa, and customized units
Unit for flow: ccm, sccm, ml/min, l/min, l/hr
Differential Pressure Leak Rate Method
Differential Pressure Leak Test
Tube Test
Medical Component Testing
EV Battery Pack Testing
IP Testing
RS-232 Communication for Data Transfer and Storage
IO connector
10 inputs and 10 outputs
Firmware Update Port
Optional Communication Protocols: Ethernet/IP, PROFINET
Scanner Integration
USB port for data download and expansion
Dimensions: 9"H x 10"W x 12"D
Weight: 12 kg
Temperature: Operating Temperature: 10 °C to 45 °C
Air Supply: Clean, dry, non-corrosive gases
Test Pressure Ranges: Vacuum, 0-1 bar, 0-7 bar, 0-10 bar, 0-16 bar Higher pressure ranges available on request